3. Leaks
Leaks are usually stopped by tightening or replacing a fitting. Be aware, however, that overtightened metal compression fittings can leak and plastic fingertights can wear out. If a fitting leak does not stop when the fitting is tightened a little, take the fitting apart and inspect for damage (e.g., distorted ferrule or particles on the sealing surface); damaged fittings should be discarded and replaced.
A. Leaky Fittings
Possible Cause |
Solution |
|
1. Loose fitting |
1. Tighten |
2. Stripped fitting |
2. Replace |
3. Overtightened fitting |
3. a. Loosen and retighten |
|
b. Replace |
4. Dirty fitting |
4. a. Disassemble and clean |
|
b. Replace |
5. Mismatched parts |
5. Use all parts from same brand |
B. Leaks at Pump
Possible Cause |
Solution |
|
1. Loose check valves |
1. a. Tighten check valve (do not over- |
|
tighten) |
|
b. Replace check valve |
2. Loose fittings |
2. Tighten fittings (do not overtighten) |
3. Mixer seal failure |
3. a. Replace mixer seal |
|
b. Replace mixer |
4. Pump seal failure |
4. Repair or replace |
5. Pressure transducer failure |
5. Repair or replace |
6. Pulse damper failure |
6. Replace pulse damper |
7. Proportioning valve failure |
7. a. Check diaphragms, replace if leaky |
|
b. Check for fitting damage, replace |
8. Purge valve |
8. a. Tighten valve |
|
b. Replace purge valve |
C. Injector Leaks
Possible Cause |
Solution |
|
1. Rotor seal failure |
1. Rebuild or replace injector |
2. Blocked loop |
2. Replace loop |
3. Loose injection-port seal |
3. Adjust |
4. Improper syringe-needle diameter |
4. Use correct syringe |
5. Waste-line siphoning |
5. Keep waste line above surface waste |
6. Waste-line blockage |
6. Replace waste line |
D. Column Leaks
Possible Cause |
Solution |
|
1. Loose endfitting |
1. Tighten endfitting |
2. Column packing in ferrule |
2. Disassemble, rinse ferrule, reassemble |
3. Improper frit thickness |
3. Use proper frit (see Frit selection guide |
|
chart) |
E. Detector Leaks
Possible Cause |
Solution |
|
1. Cell gasket failure |
1. a. Prevent excessive backpressure |
|
b. Replace gasket |
2. Cracked cell window(s) |
2. Replace window(s) |
3. Leaky fittings |
3. Tighten or replace |
4. Blocked waste line |
4. Replace waste line |
5. Blocked flow cell |
5. Rebuild or replace |
|
Previous Page Next Page
|