1. Column temperature fluctuation |
1. a. Control column and mobile phase |
(even small changes cause cyclic |
temperature |
baseline rise and fall; most |
b. Use heat exchanger before detector |
often affects refractive index |
|
and conductivity detectors, or |
|
UV detectors at high sensitivity |
|
or in direct photometric mode) |
|
2. Nonhomogeneous mobile phase |
2. a. Use HPLC-grade solvents, high-purity |
(drift usually to higher |
salts, and additives |
absorbance, rather than cyclic |
b. Degas mobile phase before use |
pattern from temperature |
c. Sparge with helium during use |
fluctuation) |
|
3. Contaminant or air buildup in |
3. a. Flush cell with methanol or other |
detector cell |
strong solvent |
|
b. If necessary, clean cell with 1 n |
|
HNO3 (never with HCl) |
4. Plugged outlet line after detector |
4. a. Unplug or replace line |
(high-pressure cracks cell |
b. Refer to detector manual to replace |
window, producing noisy baseline) |
window |
5. Mobile-phase mixing problem or |
5. a. Correct composition/flow rate |
change in flow rate |
b. To avoid, routinely monitor |
|
composition and flow rate |
6. Slow column equilibration, |
6. a. Flush with intermediate strength |
especially when changing |
solvent |
mobile phase |
b. Run 10–20 column volumes of new |
|
mobile phase before analysis |
7. Mobile phase contaminated, |
7. a. Check make-up of mobile phase |
deteriorated, or prepared from |
b. Use highest grade chemicals and HPLC |
low-quality materials |
solvents |
8. Strongly retained materials in |
8. a. Use guard column |
sample (high k) can elute as very |
b. If necessary, flush column with strong |
broad peaks and appear to be |
solvent between injections or periodi- |
a problem |
ically during analysis |
9. Mobile phase recycled but |
9. a. Reset baseline |
detector not adjusted |
b. Use new mobile phase when dynamic |
|
range of detector is exceeded |
10. Detector (UV) not set at |
10. Change wavelength to UV absorbance |
absorbance maximum but at slope |
maximum |
of curve |
|
|