.
. Proteins and Proteomics: A Laboratory ManualCSHL Press .
. . . . .
.
.
.
. .
 
.
. . .
.  FIRST VISIT?
 TRY THESE LINKS
.
. . .
.
.   Enroll for Updates
  Privacy Policy
  Purchase the book
.
.
 
 
.
. . .
.  BOOK COVER .
. . .
.
. Proteins and Proteomics: A Laboratory Manual cover .
.
CLICK TO ENLARGE
 
Buy the Book
 
   
 

4. Problems with the Chromatogram

Many problems in an LC system show up as changes in the chromatogram. Some of these can be solved by changes in the equipment; however, others require modification of the assay procedure. Selecting the proper column type and mobile phase are keys to "good chromatography."

A. Peak Tailing

Possible Cause Solution

1. Blocked frit 1. a. Reverse flush column (if allowed)
     b. Replace inlet frit
     c. Replace column
2. Column void 2. Fill void
3. Interfering peak 3. a. Use longer column
     b. Change mobile-phase and/or column/
     selectivity
4. Wrong mobile-phase pH 4. a. Adjust pH
     b. For basic compounds, a lower pH
     usually provides more symmetric peaks
5. Sample reacting with active sites 5. a. Add ion pair reagent or volatile basic

pro04-4-a
   modifier
   b. Change column

B. Peak Fronting

Possible Cause Solution

1. Low temperature 1. Increase column temperature
2. Wrong sample solvent 2. Use mobile phase for injection solvent
3. Sample overload 3. Decrease sample concentration
4. Bad column 4. See A.1. and A.2.

C. Split Peaks

Possible Cause Solution

1. Contamination on guard or 1. a. Remove guard column and attempt
   analytical column inlet    analysis

pro04-4-c
   b. Replace guard if necessary
   c. If analytical column is obstructed,
   reverse and flush
   d. If problem persists, column may be
   fouled with strongly retained
   contaminants
   e. Use appropriate restoration procedure
   f. If problem persists, inlet is probably
   plugged
   g. Change frit or replace column
2. Sample solvent incompatible with 2. Change solvent; whenever possible,
   mobile phase    inject samples in mobile phase

D. Distortion of Larger Peaks

Possible Cause Solution

1. Sample overload 1. Reduce sample size

E. Distortion of Early Peaks

Possible Cause Solution

1. Wrong injection solvent 1. a. Reduce injection volume
     b. Use weaker injection solvent

F. Tailing, Early Peaks More Than Later Ones

Possible Cause Solution

1. Extra-column effects 1. a. Replumb system (shorter, narrower
     tubing)
     b. Use smaller volume detector cell

G. Increased Tailing as k' Increases

Possible Cause Solution

1. Secondary retention effects, 1. a. Add triethylamine (basic samples)
   reversed-phase mode    b. Add acetate (acidic samples)
     c. Add salt or buffer (ionic samples)
     d. Try a different column
2. Secondary retention effects, 2. a. Add triethylamine (basic compounds)
   normal-phase mode    b. Add acetic acid
3. Secondary retention effects, 3. Add triethylamine (basic samples)
   ion-pair  

H. Acidic or Basic Peaks Tail

Possible Cause Solution

1. Inadequate buffering 1. a. Use 50–100 mm buffer concentration
     b. Use buffer with pKa equal to pH of
     mobile phase

I. Extra Peaks

Possible Cause Solution

1. Other components in sample 1. Normal
2. Late-eluting peak from previous 2. a. Increase run time or gradient slope
   injection    b. Increase flow rate
3. Vacancy or ghost peaks 3. a. Check purity of mobile phase
     b. Use mobile phase as injection solvent
     c. Reduce injection volume

J. Retention Time Drifts

Possible Cause Solution

1. Poor temperature control 1. Thermostat column
2. Mobile phase changing 2. Prevent change (evaporation, reaction,
     etc.)
3. Poor column equilibration 3. Allow more time for column equili-
     bration between runs

K. Abrupt Retention Time Changes

Possible Cause Solution

1. Flow rate change 1. Reset flow rate
2. Air bubble in pump 2. Bleed air from pump
3. Improper mobile phase 3. a. Replace with proper mobile phase
     b. Set proper mobile phase mixture on
     controller

L. Baseline Drift

Possible Cause Solution

1. Column temperature fluctuation 1. a. Control column and mobile phase
   (even small changes cause cyclic    temperature
   baseline rise and fall; most    b. Use heat exchanger before detector
   often affects refractive index  
   and conductivity detectors, or  
   UV detectors at high sensitivity  
   or in direct photometric mode)  
2. Nonhomogeneous mobile phase 2. a. Use HPLC-grade solvents, high-purity
   (drift usually to higher    salts, and additives
   absorbance, rather than cyclic    b. Degas mobile phase before use
   pattern from temperature    c. Sparge with helium during use
   fluctuation)  
3. Contaminant or air buildup in 3. a. Flush cell with methanol or other
   detector cell    strong solvent
     b. If necessary, clean cell with 1 n
     HNO3 (never with HCl)
4. Plugged outlet line after detector 4. a. Unplug or replace line
   (high-pressure cracks cell    b. Refer to detector manual to replace
   window, producing noisy baseline)    window
5. Mobile-phase mixing problem or 5. a. Correct composition/flow rate
   change in flow rate    b. To avoid, routinely monitor
     composition and flow rate
6. Slow column equilibration, 6. a. Flush with intermediate strength
   especially when changing    solvent
   mobile phase    b. Run 10–20 column volumes of new
     mobile phase before analysis
7. Mobile phase contaminated, 7. a. Check make-up of mobile phase
   deteriorated, or prepared from    b. Use highest grade chemicals and HPLC
   low-quality materials    solvents
8. Strongly retained materials in 8. a. Use guard column
   sample (high k) can elute as very    b. If necessary, flush column with strong
   broad peaks and appear to be    solvent between injections or periodi-
   a problem    ically during analysis
9. Mobile phase recycled but 9. a. Reset baseline
   detector not adjusted    b. Use new mobile phase when dynamic
     range of detector is exceeded
10. Detector (UV) not set at 10. Change wavelength to UV absorbance
   absorbance maximum but at slope    maximum
   of curve  

pro04-4-l

M. Baseline Noise (Regular)

Possible Cause Solution

1. Air in mobile phase, detector cell, 1. a. Degas mobile phase
   or pump    b. Flush system to remove air from
     detector cell or pump
2. Leak 2. a. See Section 3

pro04-4-m
   b. Check system for loose fittings
   c. Check pump for leaks, salt build-up,
   unusual noises
   d. Change pump seals if necessary
3. Incomplete mobile phase mixing 3. Mix mobile phase by hand or use less
     viscous solvent
4. Temperature effect (column at 4. Reduce differential or add head
   high temperature, detector    exchanger
   unheated)  
5. Other electronic equipment on 5. Isolate LC, detector, or recorder to
   same line    determine if source of problem is
     external; correct as necessary
6. Pump pulsations 6. Incorporate pulse dampener into system

N. Baseline Noise (Irregular)

Possible Cause Solution

1. Leak 1. a. See Section 3

pro04-4-n
   b. Check for loose fittings
   c. Check pump for leaks, salt build-up,
   unusual noises
   d. Change seals if necessary
   e. Check for detector cell leak
2. Mobile phase contaminated, 2. Check make-up of mobile phase
   deteriorated, or prepared  
   from low-quality materials  
3. Mobile phase solvents immiscible 3. Select and use only miscible solvents
4. Detector/recorder electronics 4. a. Isolate detector and recorder
     electronically
     b. Refer to instruction manual to correct
     problem
5. Air trapped in system 5. Flush system with strong solvent
6. Air bubbles in detector 6. a. Purge detector
     b. Install back-pressure device after
     detector
7. Detector cell contaminated (even 7. Clean cell by flushing with 1 n HNO3
   small amounts of contaminants    (never with HCl)
   can cause noise)  
8. Weak detector lamp 8. Replace lamp
9. Column leaking silica or packing 9. Replace column
   material  
10. Mobile phase mixer inadequate or 10. Repair or replace the mixer or mix off-
   malfunctioning    line if isocratic

O. Broad Peaks

Possible Cause Solution

1. Mobile-phase composition changed 1. Prepare new mobile phase
2. Mobile-phase flow rate too low 2. Adjust flow rate
3. Leaks (especially between column 3. a. See Section 3
   and detector)    b. Check for loose fittings
     c. Check pump for leaks, salt build-up,
     and unusual noises
     d. Change seals if necessary
4. Detector settings incorrect 4. Adjust settings
5. Extra-column effects: 5. a. Inject smaller column (e.g., 10 µl vs.
   a. Column overloaded    100 µl) or 1:10 and 1:100 dilutions of
   b. Detector response time or cell    sample
   volume too large    b. Reduce response time or use smaller
     cell
   c. Tubing between column and    c. Use as short a piece of 0.007–0.010-
   detector too long or ID    inch ID tubing as practical
   too large  
   d. Recorder response time too    d. Reduce response time
   high  
6. Buffer concentration too low 6. Increase concentration
7. Guard column contaminated/worn 7. Replace guard column
   out  
8. Column contaminated/worn out; 8. a. Replace column with new one of same
   low plate number    type
     b. If new column provides symmetrical
     peaks, flush old column with strong
     solvent
9. Void at column inlet 9. Open inlet end and fill void or replace
     column
10. Peak represents two or more 10. Change column type to improve
   poorly resolved compounds    separation
11. Column temperature too low 11. Increase temperature; do not exceed
     75°C unless higher temperatures
     are acceptable to column
     manufacturer
12. Detector time constant too large 12. Use smaller time constant

pro04-4-o

P. Loss of Resolution

Possible Cause Solution

1. Mobile phase contaminated/ 1. Prepare new mobile phase
   deteriorated (causing  
   retention time to change)  
2. Obstructed guard or analytical 2. a. Remove guard column and attempt
   column    analysis

pro04-4-p
   b. Replace guard if necessary
   c. If analytical column is obstructed,
   reverse and flush; if problem
   persists, column may be fouled
   with strongly retained contaminants
   d. Use appropriate restoration procedure;
   if problem persists, inlet is probably
   plugged
   e. Change frit or replace column

Q. All Peaks Too Small

Possible Cause Solution

1. Detector attentuation too high 1. Reduce attenuation
2. Detector time constant too large 2. Use smaller time constant
3. Injection size too small 3. Use larger sample loop
4. Improper recorder connection 4. Use correct connection

R. All Peaks Too large

Possible Cause Solution

1. Detector attenuation too low 1. Use larger attenuation
2. Injection size too large 2. Use smaller sample loop
3. Improper recorder connection 3. Use correct connection

Previous Page            Next Page

 
 
 

 
   
. .